Dry Sieving (particle-size of sand
fraction)
Objectives
Equipment and Reagents
Vibratory sieve-shaking machine with
timer
Two nests of sieves, 0.063 – 2mm at 0.5
phi intervals (see Wentworth
scale)
Stiff sieve brush
Balance with tare facility, accuracy +
0.5 g, used with a paper plate receptor.
Temperature-controlled oven, preset to
105oC + 5
Paper plates
Large sheet of thin card.
Log sheets
Methodology
-
Dry the sample(s) in the oven at 105oC
for half a day.
-
Record the total weight of sediment
sample to be sieved.
-
Ensure the nest of sieves is in the
correct order.
-
Add the sample to the top of the nest.
-
Clamp the nest into the shaker. Sieve
for 20 minutes. NOTE: Prepare a second sample for sieving with a
second nest (2-4 above). When the first nest is ready put the second
nest onto the shaker and start the next 20 minute cycle. Always have
one sample sieving while another is being weighed.
-
Weigh the first set of sieves, from the
top down. Invert each over the sheet of card, brush the mesh clean,
from both sides. Use the card as a chute to direct the sand onto the
balance receptor. Record the weight retained on each sieve. The
contents of each sieve are re-amalgamated into a single sample after
weighing.
-
Describe the nature of the material
retained on the 2mm sieves (typical and maximum dimensions, lithogenic
or biogenic origin, level of encrustation …. etc).
-
The sample is stored in a labelled
plastic bag after sieving.
-
The data from the
log sheet is entered
into an EXCEL programme, and the checksums noted and the form of the
graph checked as a quality assurance procedure.
NOTE
1: This methodology is based on British
Standard 1377.
NOTE 2: For large volume samples, the
particle-size of the gravel fraction (up to 16mm diameter) can be
determined using the same method, by adding sieves to the top of the nest.
|